Registration Dossier

Data platform availability banner - registered substances factsheets

Please be aware that this old REACH registration data factsheet is no longer maintained; it remains frozen as of 19th May 2023.

The new ECHA CHEM database has been released by ECHA, and it now contains all REACH registration data. There are more details on the transition of ECHA's published data to ECHA CHEM here.

Diss Factsheets

Physical & Chemical properties

Particle size distribution (Granulometry)

Currently viewing:

Administrative data

Link to relevant study record(s)

Reference
Endpoint:
particle size distribution (granulometry)
Type of information:
experimental study
Adequacy of study:
key study
Study period:
From 17.02.2010 to 18.02.2010
Reliability:
1 (reliable without restriction)
Rationale for reliability incl. deficiencies:
guideline study
Justification for type of information:
Test method according to OECD guideline 110 and GLP compliance.
Qualifier:
according to guideline
Guideline:
OECD Guideline 110 (Particle Size Distribution / Fibre Length and Diameter Distributions)
Deviations:
no
GLP compliance:
yes (incl. QA statement)
Type of method:
Laser scattering/diffraction
Type of distribution:
volumetric distribution
Key result
Percentile:
D50
Remarks on result:
other: 'Mass median diameter' as D50 percentile. No source field for Standard deviation.
Key result
Percentile:
D10
Mean:
102.43 µm
St. dev.:
6.541
Key result
Percentile:
D50
Mean:
263.44 µm
St. dev.:
8.392
Key result
Percentile:
D90
Mean:
590.57 µm
St. dev.:
31.254

The average and standard deviation of 5 determinations is:

X10 <= 102.43 ± 6.541 µm

X50 <= 263.44 ± 8.392 µm

X90 <= 590.57 ± 31.254 µm.

Conclusions:
The average and standard deviation of 5 determinations is:
X10 <= 102.43 ± 6.541 µm
X50 <= 263.44 ± 8.392 µm
X90 <= 590.57 ± 31.254 µm
Executive summary:

The Particle Size Distribution of the test item Ammonium Acetate was measured with Sympatec HELOS laser diffraction with wet dispersing system CUVETTE.

The average and standard deviation of 5 determinations is:

X10 <= 102.43 ± 6.541 µm

X50 <= 263.44 ± 8.392 µm

X90 <= 590.57 ± 31.254 µm.

Description of key information

Particle size distribution: Key study: The average and standard deviation of 5 determinations is:

X10 <= 102.43 ± 6.541 µm

X50 <= 263.44 ± 8.392 µm

X90 <= 590.57 ± 31.254 µm.

Test method according to OECD guideline 110 and GLP compliance.

Additional information

Key study: Experimental results: Test method according to OECD guideline 110 and GLP compliance.

The Particle Size Distribution of the test item Ammonium Acetate was measured with Sympatec HELOS laser diffraction with wet dispersing system CUVETTE.

The average and standard deviation of 5 determinations is:

X10 <= 102.43 ± 6.541 µm

X50 <= 263.44 ± 8.392 µm

X90 <= 590.57 ± 31.254 µm.